【Chemical Encyclopedia】Top Ten Issues to Pay Attention to in Liquid Chromatography HPLC
Ten Major Questions
- Dynamic Phase Unfiltered
- Failure to clean the pump in a timely manner after use
- Mobile Phase Emptying
- Is The Pressure And Flow Unstable?
- Improper Selection Of Mobile Phase For Gradient Elution
- Neglected Blank Gradient Elution
- Neglecting The Viscosity Changes Caused By Solvent Mixing
- Correct Use And Maintenance Of Six Way Valves
- Regarding The Use And Maintenance Of Chromatographic Columns
- How To Eliminate Small Probability Faults?
📝01
Dynamic Phase Unfiltered
Because dust or any other impurity particles can wear down the plunger, sealing ring, cylinder body, and one-way valve, any solid particles in the mobile phase should be removed in advance.
The mobile phase is best distilled in a glass container, and the commonly used method is filtration, which can use filters such as Millipore membranes (0.2 μ m or 0.45 μ m). The inlet of the pump should be connected to a sand filter rod (or plate). The filter of the infusion pump should be cleaned or replaced regularly.
📝02
Failure To Clean The Pump In A Timely Manner After Use
The mobile phase should not contain any corrosive substances, and the mobile phase containing buffer solution should not be retained in the pump, especially when the pump is stopped overnight or for a longer period of time.
If the mobile phase containing buffer solution is left in the pump, due to evaporation or leakage, or even just due to the static state of the solution, fine crystals of salt may precipitate, which will damage the sealing ring and plunger like the solid particles mentioned above. Therefore, it is necessary to pump pure water to thoroughly clean the pump, and then replace it with a solvent suitable for chromatographic column storage and beneficial for pump maintenance (for reverse phase bonded silica gel stationary phase, it can be methanol or methanol water).
📝03
Mobile Phase Emptying
When the pump is working, be careful to prevent the mobile phase in the solvent bottle from being used up, otherwise empty pump operation will also wear down the plunger, cylinder body or sealing ring, ultimately resulting in leakage.
What if there is no flow of mobile phase and no pressure indication?
It is possible that there is a large amount of gas inside the pump. In this case, the pressure relief valve can be opened to operate the pump at a higher flow rate (such as 5ml/min) to exhaust the bubbles. Alternatively, a 50ml syringe can be used at the pump outlet to help extract the gas. Another possible reason is that the sealing ring is worn and needs to be replaced.
📝04
Is The Pressure And Flow Unstable?
The reason may be bubbles, which need to be eliminated; Or if there are foreign objects inside the one-way valve, the one-way valve can be removed and immersed in acetone for ultrasonic cleaning. Sometimes there may be bubbles inside the sand filter rod, or it may be partially blocked by fine grains of salt or microorganisms. In this case, the sand filter rod can be removed and immersed in the mobile phase for ultrasonic removal of bubbles, or immersed in dilute acid (such as 4mol/L nitric acid) to quickly remove microorganisms, or the salt can be dissolved and immediately cleaned.
Why is the pressure too high or too low?
Perhaps the pipeline is blocked and needs to be cleared and cleaned. The reason for the decrease in pressure may be a leak in the pipeline. When checking for blockages or leaks, it should be done section by section.
When performing gradient elution, due to the mixing of multiple solvents and the constantly changing composition, some special issues arise that must be fully addressed.
PS: Gradient elution method is used for samples with complex components in liquid chromatography. In the same analysis cycle, continuously changing the concentration ratio of the mobile phase according to a certain program is called gradient elution. Thus, it is possible to achieve good separation of components with significant differences in properties in a complex sample according to their respective suitable capacity factors k.
Advantages of gradient elution:
1. Shorten the analysis cycle;
2. Improve separation capability;
3. The peak shape has been improved with little tailing;
4. Increase sensitivity. But sometimes it causes baseline drift.
📝05
Improper Selection Of Mobile Phase For Gradient Elution
Attention should be paid to the solubility of solvents, and immiscible solvents should not be used as the mobile phase for gradient elution. Some solvents are miscible within a certain proportion, but once they exceed the range, they become immiscible. Therefore, caution should be taken when using them.
When organic solvents and buffer solutions are mixed, salt crystals may also precipitate, especially when using phosphates, which requires special caution.
📝06
Neglected Blank Gradient Elution
The solvent purity required for gradient elution is higher to ensure good reproducibility.
Before conducting sample analysis, a blank gradient elution must be performed to identify solvent impurity peaks, as impurities enriched in weak solvents will be eluted by strong solvents after reaching the chromatographic column head. The solvent used for gradient elution needs to be thoroughly degassed to prevent the formation of bubbles during mixing.
📝07
Neglecting The Viscosity Changes Caused By Solvent Mixing
The viscosity of mixed solvents often varies with composition, resulting in pressure changes during gradient elution.
For example, the viscosity of methanol and water is relatively low. When the two are mixed in similar proportions, the viscosity increases significantly, and the column pressure at this time is about twice that of methanol or water as the mobile phase. Therefore, attention should be paid to preventing the pressure during gradient elution from exceeding the maximum pressure that the infusion pump or chromatographic column can withstand.
📝08
Correct Use And Maintenance Of Six Way Valves
1. The sample solution must be filtered through a 0.45 μ m filter membrane before injection to reduce particle wear on the injection valve.
2. When rotating the valve core, it should not be too slow, nor should it stay in the middle position, otherwise the flowing phase will be blocked, causing a sharp increase in pressure inside the pump, even exceeding the maximum pressure of the pump; When turning to the injection position again, excessive pressure will damage the column head.
3. To prevent buffer salt and sample residue in the injection valve, the injection valve should be flushed after each analysis. Usually, it can be rinsed with water, or first rinsed with a solvent that can dissolve the sample, and then rinsed.
📝09
Regarding The Use And Maintenance Of Chromatographic Columns
The correct use and maintenance of chromatographic columns are crucial, as even a slight mistake can reduce column efficiency, shorten service life, or even damage the column. During chromatographic operations, it is important to pay attention to the following issues in order to maintain the chromatographic column.
1. Adjust the flow rate too quickly
Avoid sudden changes in pressure and temperature, as well as any mechanical vibrations. Sudden changes in temperature or dropping the chromatographic column from a high position can affect the filling condition inside the column; The sudden increase or decrease in column pressure can also stimulate the packing inside the column, so the flow rate should be adjusted slowly, and the rotation of the valve should not be too slow during valve injection (as mentioned earlier).
2. Backwash chromatography column
Generally speaking, chromatographic columns cannot be backwashed. Only when the producer indicates that the column can be backwashed, can impurities remaining in the column head be removed by backwashing. Otherwise, recoil will quickly reduce column efficiency.
3. Pre column and protective column
Choose the appropriate mobile phase (especially pH) to avoid damaging the stationary phase. Sometimes a pre column can be connected in front of the injector. When the analysis column is bonded silica gel, the pre column is silica gel, which can make the mobile phase “saturated” with silica gel before entering the analysis column, avoiding the dissolution of the silica gel matrix in the analysis column.
To avoid injecting complex matrix samples, especially biological samples, directly into the column, pre-treatment of the samples or connecting a protective column between the injector and the chromatographic column is necessary. Protective columns are generally short columns filled with similar fixed phases. The protective column can and should be replaced frequently.
📝10
How To Eliminate Small Probability Faults?
A. Pump pressure fluctuation, with a fluctuation range of about 10 bar
resolvent:
1. The small white head was originally a bit dirty, but it has not been resolved after replacement;
After removing the valve core from the active valve, it was found that there were similar bubbles present. The valve was immersed in a methanol water solution and sonicated for 5 minutes to try to remove the bubbles. After reinstalling it and observing again, the problem was still not solved;
3. Replace the valve core with a new one, and after installation, there is no more jumping. After observing for a period of time, it is relatively stable and successfully resolved.
B. Malfunction occurred
1.plug homing over pressure;
2. There is no pressure or liquid outflow during infusion and flow rate setting;
3. The detector shutter cannot be reset. 4. After entering the high concentration sample, a peak also appears at the corresponding position in the blank.
Possible reasons:
1. The online filter is clogged;
2. The one-way valve under the pump head is blocked;
3. It could be that the detection pool is dirty, or it could be that the light path is a bit blocked. If it doesn’t work, it could be that the lamp’s lifespan has expired;
4. The small cloth is dirty and needs to be replaced.
resolvent:
1. Remove the online filter, place it in a beaker, add purified water and boil for 5-10 minutes. After that, the methanol level should be exceeded. If it still gets clogged, soak it in dilute nitric acid. If it still doesn’t work, replace it with a new one. This is a consumable in itself;
2. Remove the one-way valve under the pump head, both of which may block. When first removed, shake it vertically in your ear and you won’t hear any sound. Simply use an ear wash ball to blow the one-way valve (pay attention to the direction). If there is no response after blowing, you can use solvent ultrasound, but do not boil it in water. Generally, if it exceeds the limit and you can hear the sound by shaking it in your ear, it will not affect the experiment;
3. First, rinse the detection tank with solvent, depending on the specific sample to be tested; If it still doesn’t work after flushing, remove the detection tank and check the energy of the lamp. If the energy is still good (there are several thousand), the light path may need to be cleaned. For the first cleaning, it is recommended to have an engineer come because the component to be cleaned is located at the bottom of the detector, with a black piece and a gap in the middle, and the screws on that piece are all octagonal, making it difficult to remove;
4. Take out all 5 sample trays, and you can see a black protrusion inside (if you can’t see clearly, use a flashlight), which is where the sample needle usually draws samples. You can easily remove it by reaching in your hand, and there is a small white circle inside (which will naturally turn black if used too much). Just replace it with a new one, and usually come with 10 when buying an instrument.